This topic describes the parameters of the List / Step Sweep tab of the RF Output block.
The signal generator supports the following two types of sweeping through a set of frequency and amplitude points:
Step sweep provides a linear or logarithmic progression from one selected frequency, amplitude, or both, to another, pausing at linearly or logarithmically spaced points (steps) along the sweep.
Sweep steps are configured using the Frequency Start and Stop, Power Start and Stop, Dwell Time, and Step Points. Same Dwell Time is applied to all steps.
List sweep enables you to enter frequencies and amplitudes at unequal intervals, in nonlinear ascending, descending, or random order. Sweep steps are configured independently from the specified Frequency, Power, Dwell Time, and Waveform. Dwell time is editable at each point.
Initiate Immediate per Channel
Initiate Immediate All Channels
Edit Step Sweep (Step Sweep Configuration) Setup
Edit List Sweep (List Sweep Configuration) Setup
Enables the Sweep signaling.
GUI Location |
RF Output > List / Step Sweep > Enable Sweep |
SCPI Command |
[:SOURce][:RF<channel>]:LIST[:STATe] ON|OFF|1|0 [:SOURce][:RF<channel>]:LIST[:STATe]? |
SCPI Example |
:LIST ON :LIST? |
Dependencies |
Unavailable when Group Configuration is not Independent. To enable sweep, at least one of Frequency, Power, or Waveform Sweep State must be ON. When none of these are ON, Enable Sweep is set to OFF and is unchangeable. |
Couplings |
|
Preset |
OFF |
State Saved |
Yes |
Sets the state of frequency sweep.
GUI Location |
RF Output > List / Step Sweep > Frequency |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:FREQuency:STATe ON|OFF|1|0 [:SOURce][:RF<channel>]:LIST:FREQuency:STATe? |
SCPI Example |
:LIST:FREQ:STAT ON :LIST:FREQ:STAT? |
Couplings |
|
State Saved |
Yes |
Preset |
ON |
Sets the state of Power sweep.
GUI Location |
RF Output > List / Step Sweep > Power |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:POWer:STATe ON|OFF|1|0 [:SOURce][:RF<channel>]:LIST:POWer:STATe? |
SCPI Example |
:LIST:POW:STAT ON :LIST:POW:STAT? |
Dependencies |
Note that the Power Sweep State is not turned off when User Power Limit Enable is set to OFF.
|
State Saved |
Yes |
Preset |
ON |
Sets the state of Waveform sweep.
GUI Location |
RF Output > List / Step Sweep > Waveform |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:WAVeform:STATe ON|OFF|1|0 [:SOURce][:RF<channel>]:LIST:WAVeform:STATe? |
SCPI Example |
:LIST:WAV:STAT ON :LIST:WAV:STAT? |
Notes |
For N5185A: This setting is always set to OFF. |
Dependencies |
The Waveform Sweep State can be set to On only when the Sweep Type is set to List. |
State Saved |
Yes |
Preset |
OFF |
Couplings |
|
Sets the type of Sweep by toggling between the following two types of sweeps.
LIST - This type of sweep has arbitrary frequencies and amplitudes. Therefore, you need to specify frequency, power, dwell, and waveform for each step.
STEP - This type of sweep has equally spaced frequencies and amplitudes. Frequency and Power are calculated from step points and the same dwell time is applied to all steps.
GUI Location |
RF Output > List / Step Sweep > Sweep Type |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:TYPE STEP|LIST [:SOURce][:RF<channel>]:LIST:TYPE? |
SCPI Example |
:LIST:TYPE LIST :LIST:TYPE? |
Preset |
STEP |
Range |
Step|List |
State Saved |
Yes |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:TYPE |
Sets the direction of a List or Step sweep.
UP - This choice enables a sweep in an ascending order:
First to last point for a List sweep
Start to stop for a Step sweep
DOWN - This choice reverses the direction of the sweep.
GUI Location |
RF Output > List / Step Sweep > Sweep Direction |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:DIRection UP|DOWN [:SOURce][:RF<channel>]:LIST:DIRection? |
SCPI Example |
:LIST:DIR UP :LIST:DIR? |
Preset |
UP |
Range |
Up|Down |
State Saved |
Yes |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:DIRection |
Sets either a continuous or single List or Step sweep per channel. Execution of this command does not affect a sweep in progress.
GUI Location |
RF Output > List / Step Sweep > Sweep Repeat |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:REPeat:CONTinuous ON|OFF|1|0 [:SOURce][:RF<channel>]:LIST:REPeat:CONTinuous? |
SCPI Example |
:LIST:REP:CONT OFF :LIST:REP:CONT? |
Preset |
ON |
Range |
Single | Continuous
|
State Saved |
Yes |
Backward Compatibility SCPI |
N51xxB: :INITiate:CONTinuous[:ALL] |
Backward Compatibility Notes |
For N51xxB: Alias :INITiate:CONTinuous[:ALL] to :SOURce:RF1:LIST:REPeat:CONTinuous The original N51xxB SCPI command is categorized for all channel command since it starts from ":INITiate". The SCPI command mnemonic for N5186A/N5185A is defined to support channel 1 for the backward compatibility. |
GUI option only.
Clicking this option performs the following two tasks:
Sets the Sweep Repeat setting to Single.
Initiates a single repetition of the configured sweep.
If the trigger is set to Free Run, the sweep is initiated as soon as you arm the sweep. When the sweep is armed, the signal generator begins the sweep as soon as the trigger is received.
This option is not active if Sweep Enable is set to Off.
GUI Location |
RF Output > List / Step Sweep > Single Sweep |
SCPI Command |
No direct SCPI command is available. |
SCPI Example |
N/A |
Notes |
GUI only Momentary action No direct SCPI command is available. Run the following equivalent commands: :INITiate:CONTinuous[:ALL] OFF (when Sweep Repeat is Cont) [:SOURce][:RF<channel>]:SWEep:INITiate[:IMMediate][:ALL] [:SOURce][:RF<channel>]:TSWep |
Dependencies |
When Operation Mode is set to Manual, the Single Sweep button is disabled. |
State Saved |
No |
Configures the sweep to retrace to the first sweep point, or stop at the last sweep point upon completion of each sweep.
If set to On, the signal generator resets to the first sweep point after completion of a single sweep.
If set to Off, the signal generator stops at the last sweep point after completion of a single sweep.
GUI Location |
RF Output > List / Step Sweep > Sweep Retrace |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:RETRace ON|OFF|1|0 [:SOURce][:RF<channel>]:LIST:RETRace? |
SCPI Example |
:LIST:RETR OFF :LIST:RETR? |
Preset |
ON |
State Saved |
Yes |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:RETRace |
Sets the behavior of the Sweep Out output signal as one of the following two options/
and a single pulse for the dwell of just one selected point of the List/Sweep.
All (ALL) - the Sweep Out output signal outputs a pulse during the dwell of all points.
Single (SPOint) - the Sweep Out output signal outputs a single pulse for the dwell of just one selected point of the sweep.
GUI Location |
RF Output > List / Step Sweep > Sweep Out Points |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:SOUT ALL|SPOint [:SOURce][:RF<channel>]:LIST:SOUT? |
SCPI Example |
:LIST:SOUT ALL :LIST:SOUT? |
Preset |
ALL |
Range |
All (ALL) Single (SPOint) |
State Saved |
Yes |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:SOUT |
Enabled when Sweep Out Points is set to Single Point.
When the Sweep Out is configured to output a single pulse for the dwell of just one selected point of the List/Sweep, this setting sets that specific point of the sweep. The value is limited to the current number of points in the active columns of the list when the Sweep Out selection is Single Point.
GUI Location |
RF Output > List / Step Sweep > Sweep Out Points set to Single Point > Single Point |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:SOUT:SPOint <value> [:SOURce][:RF<channel>]:LIST:SOUT:SPOint? |
SCPI Example |
:LIST:SOUT:SPO 100 :LIST:SOUT:SPO? |
Dependencies |
|
Preset |
1 |
State Saved |
Yes |
Min |
1 |
Max |
65535 |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:SOUT:SPOint |
Sets the operating mode for the current List / Step sweep.
AUTO - This choice enables the selected sweep type to perform a sweep of all points.
MANual - This choice enables you to select a single sweep point. The selected point controls the frequency and/or amplitude according to the sweep type. Refer to the Manual Point command for selecting a sweep point.
If you set the Operation Mode to MANual, during a sweep, the sweep stops at the current point.
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep or Edit List Sweep > Operation Mode |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:MODE AUTO|MANual [:SOURce][:RF<channel>]:LIST:MODE? |
SCPI Example |
:LIST:MODE MAN :LIST:MODE? |
Preset |
AUTO |
State Saved |
Yes |
Range |
AUTO|Manual |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:LIST:MODE |
Sets the sweep point, controlling the frequency and power output when Sweep Operation Mode is Manual.
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep or Edit List Sweep > Manual Point |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:MANual <integer>|UP|DOWN [:SOURce][:RF<channel>]:LIST:MANual? |
SCPI Example |
:LIST:MAN 2 :LIST:MAN? |
Notes | If the point selected is beyond the length of the longest enabled list, then the point will be clipped and an error is generated. |
Preset |
1 |
State Saved |
Yes |
Min |
1 |
Max |
65535 (Step Sweep) 3201 (List Sweep) |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:MANual |
Remote command only.
Aborts the sweep in progress for the specified channel.
SCPI Command |
[:SOURce][:RF<channel>]:LIST:ABORt |
SCPI Example |
:LIST:ABOR |
Notes |
Momentary action |
Preset |
N/A |
State Saved |
No |
Remote command only.
Aborts the list or step sweep in progress for all channels.
If INIT:CONT[:ALL] is set to On, the sweep will re-initiate immediately.
SCPI Command |
:ABORt |
SCPI Example |
:ABOR |
Notes |
Momentary action |
Preset |
N/A |
State Saved |
No |
Backward Compatibility SCPI |
N5182B: :ABORt |
Remote command only.
Initiates the sweep on the specified channel.
SCPI Command |
[:SOURce][:RF<channel>]:LIST:INITiate[:IMMediate][:ALL] |
SCPI Example |
:LIST:INIT |
Notes |
Command is ignored if a sweep is already in progress. |
Couplings |
When Sweep Trigger Source is Immediate, a single sweep is armed and started. When Sweep Trigger Source is not Immediate , a single sweep is armed, but not started. |
State Saved |
No |
Remote command only.
Initiates the sweep on all channels.
SCPI Command |
:INITiate[:IMMediate][:ALL] |
SCPI Example |
:INIT |
Notes |
Command is ignored if a sweep is already in progress. |
Couplings |
When Sweep Trigger Source is Immediate, a single sweep is armed and started. When Sweep Trigger Source is not Immediate , a single sweep is armed, but not started. |
State Saved |
No |
Backward Compatibility SCPI |
N5182B: :INITiate[:IMMediate][:ALL] |
Remote command only.
Aborts the current sweep.
SCPI Command |
[:SOURce][:RF<channel>]:TSWeep |
SCPI Example |
:TSW |
Notes |
Momentary action |
Couplings |
When Sweep Trigger Source is Immediate, arms and starts a single sweep. When Sweep Trigger Source is not Immediate, arms a single sweep. |
State Saved |
No |
Backward Compatibility SCPI |
N5182B: [:SOURce]:TSWeep |
Remote command only.
Triggers an armed List Sweep or Step Sweep to immediately start without the occurrence of the selected trigger.
SCPI Command |
[:SOURce][:RF<channel>]:LIST:TRIGger:IMMediate |
SCPI Example |
:LIST:TRIG:IMM |
Notes |
Momentary action |
State Saved |
No |
Backward Compatibility SCPI |
N51xxB :TRIGger[:SEQuence][:IMMediate] |
Backward Compatibility Notes |
For N51xxB: Alias :TRIGger[:SEQuence][:IMMediate] to :SOURce:RF1:LIST:TRIGger:IMMediate |
The Edit Step Sweep button opens the Step Sweep screen that allows you to configure the steps for Step Sweep. You can configure a minimum of 2 steps and maximum 65535 steps for Step sweep.
Sets the first frequency point in a Step sweep.
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep > Frequency Start |
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:FREQuency:STARt <freq> [:SOURce][:RF<channel>]:SWEep:FREQuency:STARt? |
SCPI Example |
:SWE:FREQ:STAR 1e9 :SWE:FREQ:STAR? |
Preset |
Maximum RF Frequency (CW). See Frequency CW |
State Saved |
Yes |
Min | Minimum RF Frequency (CW). See Frequency CW |
Max | Maximum RF Frequency (CW). See Frequency CW |
Resolution |
0.00001 Hz |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:FREQuency:STARt |
Backwards Compatibility Notes |
For N51xxB: Alias [:SOURce][:RF<channel>]:FREQuency:STARt to [:SOURce][:RF<channel>]:SWEep:FREQuency:STARt |
Sets the last frequency point in the Step sweep.
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep > Frequency Stop |
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:FREQuency:STOP <freq> [:SOURce][:RF<channel>]:SWEep:FREQuency:STOP? |
SCPI Example |
:SWE:FREQ:STOP 3e9 :SWE:FREQ:STOP? |
Preset |
Maximum RF Frequency (CW). See Frequency CW |
State Saved |
Yes |
Min | Minimum RF Frequency (CW). See Frequency CW |
Max | Maximum RF Frequency (CW). See Frequency CW |
Resolution |
0.00001 Hz |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:FREQuency:STOPt |
Backward Compatibility Notes |
For N51xxB: Alias [:SOURce][:RF<channel>]:FREQuency:STOPt to [:SOURce][:RF<channel>]:SWEep:FREQuency:STOP |
Remote command only
Sets the center frequency for a Step Sweep. The center frequency symmetrically divides the selected frequency span and is coupled to the start and stop frequency settings. The frequency range and reset values are dependent on the signal generator model and option number.
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:FREQuency:CENTer <freq> [:SOURce][:RF<channel>]:SWEep:FREQuency:CENTer? |
SCPI Example |
:SWE:FREQ:CENT 3 GHZ :SWE:FREQ:CENT? |
Couplings | Frequency Start and Frequency Stop are calculated with Frequency Center and Frequency Span. Changing Frequency Center alters Frequency Start and Frequency Stop. |
Preset |
Maximum RF Frequency (CW). See Frequency CW |
State Saved |
Yes |
Min | Minimum RF Frequency (CW). See Frequency CW |
Max | Maximum RF Frequency (CW). See Frequency CW |
Resolution |
1 Hz |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:FREQuency:CENTer |
Backwards Compatibility Notes |
For N51xxB: Alias [:SOURce][:RF<channel>]:FREQuency:CENTer to [:SOURce][:RF<channel>]:SWEep:FREQuency:CENTer |
Remote command only
Sets the length of the step sweep frequency range for the specified channel. Span setting is symmetrically divided by the selected center frequency and is coupled to the start and stop frequency settings.
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:FREQuency:SPAN <freq> [:SOURce][:RF<channel>]:SWEep:FREQuency:SPAN? |
SCPI Example |
:SWE:FREQ:SPAN 3 GHZ :SWE:FREQ:SPAN? |
Notes | Span value returns a negative value when Frequency Start is bigger than Frequency Stop. The query for the optional MAXimum or MINimum returns the span frequency limits for the current Frequency Center. |
Dependencies |
Changing Frequency Span alters Frequency Start and Frequency Stop settings. |
Preset |
0 Hz |
State Saved |
Yes |
Min |
- (maximum RF Frequency (CW) - minimum RF Frequency (CW)) See Frequency CW |
Max |
(maximum RF Frequency (CW) - minimum RF Frequency (CW)) See Frequency CW |
Resolution |
1 Hz |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:FREQuency:SPAN |
Backward Compatibility Notes |
For N51xxB: Alias [:SOURce][:RF<channel>]:FREQuency:SPAN to [:SOURce][:RF<channel>]:SWEep:FREQuency:SPAN |
Sets the first amplitude point in a Step sweep.
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep > Power Start |
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:POWer:STARt <value><unit> [:SOURce][:RF<channel>]:SWEep:POWer:STARt? |
SCPI Example |
:SWE:POW:STAR -60 :SWE:POW:STAR? |
Preset |
Minimum RF Power |
State Saved |
Yes |
Min |
Minimum RF Power |
Maximum |
Maximum RF Power |
Resolution |
0.01 Bm |
Backwards Compatibility SCPI |
N51xxB: [:SOURce]:POWer:STARt |
Backwards Compatibility Notes |
For N51xxB: Alias [:SOURce][:RF<channel>]:POWer:STARt to [:SOURce][:RF<channel>]:SWEep:POWer:STARt |
Sets the last amplitude point in a Step sweep.
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep > Power Stop |
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:POWer:STOP <value><unit> [:SOURce][:RF<channel>]:SWEep:POWer:STOP? |
SCPI Example |
:SWE:POW:STOP -20 :SWE:POW:STOP? |
Preset |
Minimum RF Power |
State Saved |
Yes |
Min |
Minimum RF Power |
Max |
Maximum RF Power |
Resolution |
0.01 Bm |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:POWer:STOP |
Backward Compatibility Notes |
For N51xxB: Alias :[SOURce][:RF<channel>]:POWer:STOP to [:SOURce][:RF<channel>]:SWEep:POWer:STOP |
Sets the number of step sweep points.
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep > Number of Points |
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:POINts <integer> [:SOURce][:RF<channel>]:SWEep:POINts? |
SCPI Example |
:SWE:POIN 201 :SWE:POIN? |
Preset |
101 |
State Saved |
Yes |
Min |
2 |
Max |
65535 |
Resolution |
1 |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:SWEep:POINts |
Query only. Returns the current sweep point.
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:CPOint? |
SCPI Example |
:SWE:CPO? |
Notes |
When Sweep Type is Step, returns the current sweep point of Step Sweep. When Sweep Type is List, returns the current sweep point of List Sweep. |
State Saved |
No |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:SWEep:CPOint |
Sets the dwell time for a Step sweep. The dwell time is the amount of time the sweep is guaranteed to pause after setting the frequency and/or power for the current point.
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep > Step Dwell Time |
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:DWELl <time> [:SOURce][:RF<channel>]:SWEep:DWELl? |
SCPI Example |
:SWE:DWEL 0.01 :SWE:DWEL? |
Preset |
2.00e-3 |
State Saved |
Yes |
Min |
100 us |
Max |
100 s |
Resolution |
1e-6 |
Backward Compatibility SCPI |
N5182B: [:SOURce]:SWEep:DWELl |
Sets the sweep mode. The instrument uses the specified start frequency, stop frequency, and number of points for both linear and log sweeps.
LINear: The difference between frequency points is constant.
LOGarithmic: The ratio between frequency points is constant.
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep > Step Spacing |
SCPI Command |
[:SOURce][:RF<channel>]:SWEep:SPACing LINear|LOGarithmic [:SOURce][:RF<channel>]:SWEep:SPACing? |
SCPI Example |
:SWE:SPAC LOG :SWE:SPAC? |
Preset |
LIN |
Range |
Linear (LIN) Logarithmic (LOG) |
State Saved |
Yes |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:SWEep:SPACing |
Sets the step size for a linear step sweep in frequency (difference between frequency points).
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep > Linear Frequency Step |
SCPI Command |
[:SOURce][:RF<channel>]:SWEep[:FREQuency]:STEP[:LINear <value><unit> [:SOURce][:RF<channel>]:SWEep[:FREQuency]:STEP[:LINear? |
SCPI Example |
:SWE:STEP 10e6 :SWE:STEP? |
Notes |
Linear frequency step size returns a negative value when Frequency Start is bigger than Frequency Stop. |
Couplings | Setting the Linear Frequency Step Size will determine the number of points in the step sweep, based on the current start and stop frequencies. Due to the integer number of step points, the step size may be adjusted to yield a true linear sweep between the start and stop frequencies. |
Preset |
0.0 Hz |
State Saved |
Yes |
Min |
|
Max |
|
Resolution |
0.00001 Hz |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:SWEep[:FREQuency]:STEP:[LINear] |
This setting is displayed when Step Spacing is set to Logarithmic.
Sets the step size for a logarithmic step sweep in frequency (ratio between frequency points), expressed as a percentage (%).
GUI Location |
RF Output > List / Step Sweep > Edit Step Sweep > Logarithmic Frequency Step |
SCPI Command |
[:SOURce][:RF<channel>]:SWEep[:FREQuency]:STEP:LOGarithmic <value> [:SOURce][:RF<channel>]:SWEep[:FREQuency]:STEP:LOGarithmic? |
SCPI Example |
:SWE:STEP:LOG 10 :SWE:STEP:LOG? |
Couplings | Setting the logarithmic frequency step size will determine the number of points in the step sweep, based on the current start and stop frequencies. Due to the integer number of step points, the step size may be adjusted to yield a true logarithmic sweep between the start and stop frequencies. |
Notes |
This setting is indicated when Step Spacing is Logarithmic. |
Preset |
0.0 |
State Saved |
Yes |
Min |
0.0 |
Max |
- |
Resolution |
0.01 |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:SWEep[:FREQuency]:STEP:LOGarithmic |
The Edit List Sweep button opens the List Sweep screen that allows you to configure the steps for List Sweep. You can configure a minimum of 2 steps and maximum 3201 steps for List sweep.
The table row sizes of Frequency, Power, Waveform and Dwell Time are always same.
Sets the frequency values for the current List sweep points. The maximum number of List sweep points is 3,201. The variable is expressed in units of hertz (Hz). The setting enabled by this command is not affected by signal generator power–on, preset, or *RST.
GUI Location |
RF Output > List / Step Sweep > Edit List Sweep > Frequency |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:FREQuency <value>,<value> [:SOURce][:RF<channel>]:LIST:FREQuency? |
SCPI Example |
:LIST:FREQ 1e9,2e9 :LIST:FREQ? |
Dependencies |
The list size is adjusted so that the same number is used in Frequency, Power, Waveform, and Dwell Time. The list number is the maximum row size (list size) in Frequency, Power, Waveform and Dwell Time. |
Notes |
|
Preset |
Maximum RF Frequency (CW) See Frequency CW |
State Saved |
Yes |
Min |
Minimum RF Frequency (CW) See Frequency CW |
Max |
Maximum RF Frequency (CW) See Frequency CW |
Resolution |
0.00001 Hz |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:LIST:FREQuency. |
Sets the amplitude for the current List sweep points.
GUI Location |
RF Output > List / Step Sweep > Edit List Sweep > Power |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:POWer <ampl>,<ampl> [:SOURce][:RF<channel>]:LIST:POWer? |
SCPI Example |
:LIST:POW -50,-40 :LIST:POW? |
Dependencies |
The list size is adjusted so that the same number is used in Frequency, Power, Waveform and Dwell Time. The list number is the maximum row size (list size) in Frequency, Power, Waveform and Dwell Time. |
Notes |
|
Preset |
Minimum RF Power |
State Saved |
Yes |
Min |
Minimum RF Power |
Max |
Maximum RF Power |
Resolution |
0.01 dBm |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:POWer |
Sets the dwell time for the current List sweep points. Dwell time is the amount of time the sweep is guaranteed to pause after setting the frequency and/or power for the current point, expressed in seconds. It is used when the trigger source of the Point trigger is set to IMMediate.
GUI Location |
RF Output > List / Step Sweep > Edit List Sweep > Dwell Time |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:DWELl <time>,<time> [:SOURce][:RF<channel>]:LIST:DWELl? |
SCPI Example |
:LIST:DWEL 0.001,0.002 :LIST:DWEL? |
Dependencies |
The list size is adjusted so that the same number is used in Frequency, Power, Waveform and Dwell Time. The list number is the maximum row size (list size) in Frequency, Power, Waveform and Dwell Time. |
Notes |
|
Preset |
2.00e-3 |
State Saved |
Yes |
Min |
0.0001 |
Max |
100 |
Resolution |
1e-6 (1 us) |
Backwards Compatibility SCPI |
N51xxB: [:SOURce]:LIST:DWELl |
Sets Dwell Type of the list sweep dwell time.
List: This choice selects the dwell times from the list sweep.Refer to the Dwell Time command for setting the dwell time for List sweep points.
Step: This choice selects the dwell time from the step sweep.Refer to the Step Dwell Time command for setting the dwell time for Step sweep.
SCPI Command |
[:SOURce][:RF<channel>]:LIST:DWELl:TYPE LIST|STEP [:SOURce][:RF<channel>]:LIST:DWELl:TYPE? |
SCPI Example |
:LIST:DWEL:TYPE STEP :LIST:DWEL:TYPE? |
Couplings |
When set to List, Dwell Type List State is set to ON. When set to Step, Dwell Type List State is set to OFF. |
Preset |
LIST |
State Saved |
Yes |
Range |
List | Step |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:DWELl:TYPE |
Sets the state of the dwell time for the list sweep points between the values defined in the list sweep and the value for the step sweep.
1|ON - This choice selects the dwell times from the List sweep. Refer to the Dwell Time command for setting the List dwell time.
0|OFF - This choice selects the dwell time from the Step sweep. Refer to the Step Dwell Time command for setting the Step dwell time.
GUI Location |
RF Output > List / Step Sweep > Edit List Sweep > Dwell Time checkbox |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:DWELl:TYPE:LIST:STATe ON|OFF|1|0 [:SOURce][:RF<channel>]:LIST:DWELl:TYPE:LIST:STATe? |
SCPI Example |
:LIST:DWEL:TYPE:LIST:STAT ON :LIST:DWEL:TYPE:LIST:STAT? |
Couplings |
When set to ON, Dwell Type is set to List. When set to OFF, Dwell Type is set to Step. |
Preset |
ON |
State Saved |
Yes |
Remote command only.
Query only. Returns the current sweep point.
SCPI Command |
[:SOURce][:RF<channel>]:LIST:CPOint? |
SCPI Example |
:LIST:CPO? |
Notes |
When Sweep Type is Step, returns the current sweep point of Step Sweep. When Sweep Type is List, returns the current sweep point of List Sweep. |
State Saved |
No |
Backwards Compatibility SCPI |
N5182B: [:SOURce]:LIST:CPOint |
Sets the waveform values for the current List waveform sequence.
GUI Location |
RF Output > List / Step Sweep > Edit List Sweep > Waveform checkbox selected > Select |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:WAVeform <string>,<string> [:SOURce][:RF<channel>]:LIST:WAVeform? |
SCPI Example |
:LIST:WAV "","WFM1:TEST" :LIST:WAV? |
Dependencies | The list size is adjusted so that the same number is used in Frequency, Power, Waveform and Dwell Time. The list number is the maximum row size (list size) in Frequency, Power, Waveform and Dwell Time. |
Notes |
CW is represented by "" The number of lists: Min 1, Max 3201 elements The header information (Sample rate, RMS, Scale, Occupied Bandwidth) is required for the waveform of list sweep. Waveform files which have header information are accepted for the waveform selection. Files on ARB memory cannot be selected for the list sweep. |
Preset |
"" |
State Saved |
Yes |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:WAVeform |
Remote command only.
Query only. Returns the current list sweep file for the number of frequency points.
SCPI Command |
[:SOURce][:RF<channel>]:LIST:FREQuency:POINts? |
SCPI Example |
:LIST:FREQ:POIN? |
State Saved |
No |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:FREQuency:POINts? |
Remote command only.
Query only. Returns the number of power points in the current list sweep file.
SCPI Command |
[:SOURce][:RF<channel>]:LIST:POWer:POINts? |
SCPI Example |
:LIST:POW:POIN? |
State Saved |
No |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:POWer:POINts? |
Remote command only.
Query only. Returns the signal generator for the number of waveform points in the current List sweep file.
SCPI Command |
[:SOURce][:RF<channel>]:LIST:WAVeform:POINts? |
SCPI Example |
:LIST:WAV:POIN? |
State Saved |
No |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:WAVeform:POINts? |
Remote command only.
Query only. Returns the signal generator for the number of dwell time points in the current List sweep file.
SCPI Command |
[:SOURce][:RF<channel>]:LIST:DWELl:POINts? |
SCPI Example |
:LIST:DWEL:POIN? |
State Saved |
No |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:DWELl:POINts? |
Imports a List sweep table file.
Supported file formats:
Binary - This format is compatible with MXG-B stored List sweep files.
CSV - New format supported for N5186A.
GUI Location |
RF Output > List / Step Sweep > Edit List Sweep > Import/Export/Preset > Import List Sweep Table |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:LOAD "<file name>" |
SCPI Example |
:LIST:LOAD "LIST1" |
Notes |
Momentary action |
State Saved |
No |
Backward Compatibility SCPI |
N51xxB: :MEMOry:LOAD:LIST |
Backwards Compatibility Notes |
For N51xxB: Alias :MEMOry:LOAD:LIST to :SOURce:RF1:LIST:LOAD |
Stores the current list sweep table data to a CSV file.
GUI Location |
RF Output > List / Step Sweep > Edit List Sweep > Import/Export/Preset > Export List Sweep Table |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:STORe "<file name>" |
SCPI Example |
:LIST:STOR "LIST1" |
Notes |
Momentary action |
State Saved |
No |
Replaces the current list sweep data with the default setting consisting of one point at a frequency, amplitude, and dwell time.
GUI Location |
RF Output > List / Step Sweep > Edit List Sweep > Import/Export/Preset > Preset List with Defaults |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:TYPE:LIST:INITialize:PRESet |
SCPI Example |
:LIST:TYPE:LIST:INIT:PRES |
State Saved |
No |
Backward Compatibility SCPI |
N5182B: [:SOURce]:LIST:TYPE:LIST:INITialize:PRESet |
Replaces the loaded list sweep data with the settings from the current step sweep data points.
GUI Location |
RF Output > List / Step Sweep > Edit List Sweep > Import/Export/Preset/ Preset List with Step Sweep |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:TYPE:LIST:INITialize:FSTep |
SCPI Example |
:LIST:TYPE:LIST:INIT:FST |
Notes | The maximum number of List sweep points is 3,201. When copying the Step sweep settings to a List sweep, ensure that the number of points in the Step sweep do not exceed the maximum List sweep points. |
State Saved |
No |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:LIST:TYPE:LIST:INITialize:FSTep |
Click this button to access the List/Step Sweep Trigger screen. You can use this screen to configure the sweep trigger for List Sweep and Step Sweep.
This section describes the settings of Sweep Trigger.
Sets the sweep trigger type.
Free Run - Immediately and continuously triggers the sweep.
Single - Triggers the sweep for one sweep when the sweep trigger source is applied.
Trigger & Run - The sweep starts when the sweep trigger source is applied and then continues to sweep.
GUI Location |
RF Output > List / Step Sweep > Edit Triggers > Sweep Trigger > Trigger Type |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:SWEep:TRIGger:TYPE FRUN|SINGle|TRUN [:SOURce][:RF<channel>]:LIST:SWEep:TRIGger:TYPE? |
SCPI Example |
:LIST:SWE:TRIG:TYPE SING :LIST:SWE:TRIG:TYPE? |
Dependencies | When you set the Trigger Type to Single or Trigger & Run, the Trigger Source is automatically set to Key if it is currently set to Immediate. |
Couplings |
When Trigger Source is set to Immediate, Trigger Type is set to FreeRun (if it is currently not set to FreeRun). |
Preset |
FRUN |
Range |
|
State Saved |
Yes |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:LIST:SWEep:TRIGger:TYPE |
Per-channel setting
Sets the source for triggering a sweep. Following options are available.
KEY: Sweep runs once when the Trigger key on the front panel is pressed.
BUS: Sweep runs once per SCPI *TRG command or GET (Group Execute Trigger).
EXTernal: Selects the use of an external trigger source.
IMMediate: Enables immediate triggering of the sweep event.
GTRigger - This choice enables the global trigger. Requires Option PCH.
GUI Location |
RF Output > List / Step Sweep > Edit Triggers > Sweep Trigger > Trigger Source |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:SWEep:TRIGger:SOURce KEY|BUS|EXTernal|IMMediate|GTRigger [:SOURce][:RF<channel>]:LIST:SWEep:TRIGger:SOURce? |
SCPI Example |
:LIST:SWE:TRIG:SOUR EXT :LIST:SWE:TRIG:SOUR? |
Notes |
The wait for the BUS, EXTernal, or KEY trigger can be bypassed by sending the Trigger Immediate command. |
Couplings |
When you set the Trigger Type to Single or Trigger & Run, the Trigger Source is automatically set to Key if it is currently set to Immediate. |
Preset |
IMM |
Range |
Key (KEY) Bus (BUS) External (EXTernal) Immediate (IMMediate) Global Trigger (GTRigger) |
State Saved |
Yes |
Backward Compatibility SCPI |
N51xxB: :TRIGger[:SEQuence]:SOURce |
Backward Compatibility Notes |
For N51xxB: Alias :TRIGger[:SEQuence]:SOURce to :SOURce:RF1:LIST:SWEep:TRIGger:SOURce The selection MANual works as same as KEY. The selection INTernal is not supported in this release. |
Sets the external trigger source of Sweep Trigger.
GUI Location |
RF Output > List / Step Sweep > Edit Triggers > Sweep Trigger > Trigger Source set to External > Trigger Source |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:SWEep:TRIGger:EXTernal:SOURce <integer> [:SOURce][:RF<channel>]:LIST:SWEep:TRIGger:EXTernal:SOURce? |
SCPI Example |
:LIST:SWE:TRIG:EXT:SOUR 2 :LIST:SWE:TRIG:EXT:SOUR? |
Couplings | If the specified connector was being used (not None) as an output, then the Trigger Output Routing of that connector is set to None and a Settings Conflict error message is generated. |
Preset |
1 |
State Saved |
Yes |
Min |
1 |
Max |
6 (N5185A / N5186A) |
Backward Compatibility SCPI |
N51xxB: :TRIGger:EXTernal:SOURce |
Backwards Compatibility Notes |
For N51xxB: Alias :TRIGger:EXTernal:SOURce to :SOURce:RF1:LIST:SWEep:TRIGger:EXTernal:SOURce For N5186A,/ N5185A:
|
Remote command only
Sets the polarity of the external trigger 1 of channel 1.
POSitive - The signal generator triggers an event when it detects a rising edge on the source signal.
NEGative - The signal generator triggers an event when it detects a falling edge on the source signal.
Refer to the Trigger Input Slope command.
Preset |
POS |
Range |
Positive|Negative |
State Saved |
Yes |
Backwards Compatibility SCPI |
N51xxB: :TRIGger[:SEQuence]:SLOPe |
Backwards Compatibility Notes |
For N51xxB: Alias :TRIGger[:SEQuence]:SLOPe to :ROUTe:CONNectors:RF1:TRIG[1]:INPut:SLOPe |
This section describes the settings of Point Trigger for a List or Step sweep event.
Sets the point trigger source for a List or Step sweep event.
BUS: Sweep point runs once per SCPI *TRG command or GET (Group Exexute Trigger)
EXTernal - Selects the use of an external trigger source.
IMMediate - Enables immediate triggering of the sweep point event.
GTRigger - Requires Option PCH.
GUI Location |
RF Output > List / Step Sweep > Edit Triggers > Point Trigger > Trigger Source |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:TRIGger:SOURce KEY|BUS|EXTernal|IMMediate|GTRigger [:SOURce][:RF<channel>]:LIST:TRIGger:SOURce? |
SCPI Example |
:LIST:TRIG:SOUR EXT :LIST:TRIG:SOUR? |
Preset |
IMMediate |
Range |
Key|Bus|External|Immediate|Global Trigger |
State Saved |
Yes |
Backwards Compatibility SCPI |
N51xxB: [:SOURce]:LIST:TRIGger:SOURce |
Backwards Compatibility Notes |
For N51xxB: The selection MANual works as same as KEY. The selection INTernal is not supported on this release. |
Sets the external trigger source of Point Trigger.
GUI Location |
RF Output > List / Step Sweep > Edit Triggers > Point Trigger > Trigger Source set to External > Trigger Source |
SCPI Command |
[:SOURce][:RF<channel>]:LIST:TRIGger:EXTernal:SOURce <integer> [:SOURce][:RF<channel>]:LIST:TRIGger:EXTernal:SOURce? |
SCPI Example |
:LIST:TRIG:EXT:SOUR 2 :LIST:TRIG:EXT:SOUR? |
Couplings | If the specified connector was being used (not None) as an output, then the Trigger Output Routing of that connector is set to None and a Settings Conflict error message is generated. |
Preset |
1 |
State Saved |
Yes |
Min |
1 |
Max |
6 (N5185A / N5186A) |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:LIST:TRIGger:EXTernal:SOURce |
Backwards Compatibility Notes |
For N51xxB: For N5186A, N5185A:
|
Remote command only
Sets the polarity of the external trigger 1 .
POSitive - The signal generator triggers an event when it detects a rising edge on the source signal.
NEGative - The signal generator triggers an event when it detects a falling edge on the source signal.
Refer to the Trigger Input Slope command.
Preset |
POS |
Range |
Positive|Negative |
State Saved |
Yes |
Backward Compatibility SCPI |
N51xxB: [:SOURce]:LIST:TRIGger:SLOPe |